Doubly tuned solenoidal resonators for small animal imaging and spectroscopy at 1.5 Tesla.

Publication Type Academic Article
Authors Ballon D, Graham M, Miodownik S, Koutcher J
Journal Magn Reson Imaging
Volume 7
Issue 2
Pagination 155-62
Date Published 01/01/1989
ISSN 0730-725X
Keywords Animals, Laboratory, Magnetic Resonance Imaging, Magnetic Resonance Spectroscopy
Abstract The design and construction of solenoidal resonators for use with small animals in a 1.5-Tesla clinical imaging system are described. The coils have been designed to exploit the B1 distributions of two resonant modes of a four-turn solenoid whose windings are in parallel. Both singly and doubly tuned versions have been constructed. 1H images of normal and pathologic anatomy in mice and rats as well as a 31P spectrum of a Walker 256 rat sarcoma are presented. A primary advantage of this design is that the coils are easy to build and implement while providing the necessary sensitivity to allow high quality images to be obtained with no changes to the hardware or software of the clinical unit.
DOI 10.1016/0730-725x(89)90699-1
PubMed ID 2716483
Back to Top