Sensitivity-enhanced echo-planar MRI at 1.5T using a 5 x 5 mesh dome resonator.

Publication Type Academic Article
Authors Meyer K, Kim K, Li T, Tulipano P, Lee K, DeLaPaz R, Hirsch J, Ballon D
Journal Magn Reson Med
Volume 36
Issue 4
Pagination 606-12
Date Published 10/01/1996
ISSN 0740-3194
Keywords Echo-Planar Imaging
Abstract In this work a 5 x 5 mesh dome resonator that has been optimized for functional brain imaging is presented. The resonator was reduced in length and diameter compared with previous versions to reduce sample losses, thus enhancing the signal-to-noise ratio of the acquired data. In addition, a 5 x 5 mesh design was employed, which offered improved axial homogeneity over an earlier 3 x 3 mesh version. The new resonator exhibited high sensitivity and good homogeneity over the brain volume, permitting analysis of functional activation over large areas of the cerebral cortex. In a direct comparison with a standard clinical head-imaging resonator, the high sensitivity of the 5 x 5 mesh dome resonator resulted in greater statistical confidence in functional activation.
DOI 10.1002/mrm.1910360416
PubMed ID 8892214
Back to Top