Publication Type Academic Article
Authors Saniour I, Robb F, Taracila V, Mishra V, Vincent J, Voss H, Kaplitt M, Chazen J, Winkler S
Journal IEEE Access
Volume 10
Pagination 25062-25072
Date Published 02/25/2022
ISSN 2169-3536
Abstract Magnetic resonance guided focused ultrasound (MRgFUS) is a non-invasive therapeutic modality for neurodegenerative diseases that employs real-time imaging and thermometry monitoring of targeted regions. MRI is used in guidance of ultrasound treatment; however, the MR image quality in current clinical applications is poor when using the vendor built-in body coil. We present an 8-channel, ultra-thin, flexible, and acoustically transparent receive-only head coil design (FUS-Flex) to improve the signal-to-noise ratio (SNR) and thus the quality of MR images during MRgFUS procedures. Acoustic simulations/experiments exhibit transparency of the FUS-Flex coil as high as 97% at 650 kHz. Electromagnetic simulations show a SNR increase of 13× over the body coil. In vivo results show an increase of the SNR over the body coil by a factor of 7.3 with 2× acceleration (equivalent to 11× without acceleration) in the brain of a healthy volunteer, which agrees well with simulation. These preliminary results show that the use of a FUS-Flex coil in MRgFUS surgery can increase MR image quality, which could yield improved focal precision, real-time intraprocedural anatomical imaging, and real-time 3D thermometry mapping.
DOI 10.1109/access.2022.3154824
PubMed ID 35600672
PubMed Central ID PMC9119199